The resolution when the glove box encounters these faults

The resolution when the glove box encounters these faults


one. Huge and modest warehouses can not replenish gas

Reasons: ① The air gap in the glove box leading to the large and small warehouse might be blocked by plastic movie, etc . ② Whether or not the solenoid valve is not open up.

Approach: ① Decide the place of the air hole and very clear the pipeline according to the pipeline path ② Check out the circuit and solenoid valve.

2. The huge and small warehouses are not able to be pumped

Motives: ① The suction holes in the big and tiny bins are blocked by plastic film, etc. or the doorway is leaking ② The bin doorways are not shut tightly, or the seal ring of the big and little bins is unfastened.

Protector Combination Glove Boxes Protector Combination Glove Boxes Protector Combination Glove Boxes : ① Figure out the place of the air gap and obvious the pipeline in accordance to the route of the pipeline ② Near the warehouse door and check out the dimensions of the sealing ring ③ Check out whether the doing work air force is also lower (because the huge warehouse exhaust valve is a pneumatic valve).

three. The box force continues to increase throughout cleansing

Cause: The cleaning valve is not opened, the consumption pressure is also high, and the pipeline is not clean.

Techniques: ①Ensure that the pipeline is smooth, and it is best not to bend it ②When cleansing, the relative strain in the tank is continuously beneath .5 mbar.

four. The primary valve of the purification column is faulty

Causes: ① The operating gas force is much less than .5Mpa ② The primary valve sign opinions failure.

Approaches: ①Replace the doing work gas cylinder, or verify the stress of the auxiliary gauge of the stress minimizing valve at .5Mpa

five. The cycle are not able to be opened

Cause: Lower-voltage or higher-voltage protection of fan inverter.

Approach: Right after the voltage is steady, shut down and restart following one moment.